Does a flat piece of contaminated steel plate meet a release
Does a flat piece of contaminated steel plate meet a release criterion of 1000 dpm/100 cm2 if a 15 cm2 probe with an efficiency of 10%, in contact with the surface, gives 200 net counts per minute? Justify your answer.
Solution
number of counts decays per minute is
N = ( 1000 decays/min)/ (100 cm2 ))( 15 cm2 )
= 150 counts/min
since detector has efficiency 10%
N =15 counts /min